# CPD User Guide
|
|
## Changing the sample
|
|
1. Stop the turbopump:
|
|
1. Press the stop button on the control panel
|
|
2. Close the valve between pump and chamber (on top of the turbo pump), by screwing it into the pump
|
|
2. Ventilate the chamber
|
|
1. Open the gate valve
|
|
2. Carefully open the needle valve (usually less than 1/8 turn)
|
|
3. When the chamber can be moved, close both valves again
|
|
3. Take off the chamber
|
|
1. Make some space to put the chamber (mind the grease on the bottom edge)
|
|
2. Carefully lift the chamber. Good places to grab it are the tubing on the left and the right bottom edge.
|
|
***Caution***: The tube connecting the chamber with the pump will exert a force to the right!
|
|
**Make sure you do not crash the chamber into the sample tray!**
|
|
4. Move the sample away from the gold mesh
|
|
1. Move the tray fully down by turning the upper micrometer screw *counterclockwise*
|
|
2. Move the tray fully to the left using the larger micrometer screw
|
|
3. Loosen the screw on the clamp holding the sample in place and remove the sample
|
|
5. Prepare the new sample
|
|
1. Put the sample on the tray and place the clamp on it.
|
|
2. Tighten the clamp screw (you might need to grab the nut on the bottom side)
|
|
3. Move the tray in x-position using the large micrometer screw.
|
|
The gold mesh should be well aligned with the sample surface
|
|
4. Carefully move the tray up by turning the small micrometer screw *clockwise*.
|
|
Leave about half a millimeter of space between the sample and the gold mesh
|
|
**Do not crash the sample into the gold mesh!**
|
|
6. Put the chamber back on
|
|
1. Put the chamber back on and align it on the two posts on the bottom and right
|
|
2. Check the backside to make sure that no cables are sticking out
|
|
7. Vacuum the chamber
|
|
1. Make sure the nitrogen venting valves are closed
|
|
2. **Make sure the turbopump is not spinning anymore**
|
|
3. When the turbopump has fully stopped, very slowly open the valve on top of the pump
|
|
4. Fully open the valve **TODO: hier noch irgendwelche checks?**
|
|
5. **Wait** until the vacuum is in the lower $10^{-2}$ range (read the value from the sensor attached to the turbopump)
|
|
6. Only when this target pressure is reached, turn on the turbopump (control panel on top)
|
|
7. Wait until the desired vacuum for measurement is reached (usually $10^{-5}$)
|
|
- $10^{-4}$: **TODO**
|
|
- $10^{-5}$: **TODO**
|
|
- $10^{-6}$: after 1-2 days
|
|
|
|
If the pressure does not drop enough: Use isopropanol to inspect for leaks
|
|
|
|
## CPD Controller Calibration
|
|
**TODO**
|
|
|
|
This process is not always necessary.
|
|
1. Set CPD Controller to Manual, Offset=Gold Work Function, Filter off (so you can see what happens immediately), Oscillator to ~2
|
|
2. Move the micrometer screw close until a signal appears
|
|
3. Set Automatic offset so that the output is 0V
|
|
Calibration: Fresh graphite probe
|
|
|
|