Changing the sample
- Stop the turbopump:
- Press the stop button on the control panel
- Close the valve between pump and chamber (on top of the turbo pump), by screwing it into the pump
- Ventilate the chamber
- Open the gate valve
- Carefully open the needle valve (usually less than 1/8 turn)
- When the chamber can be moved, close both valves again
- Take off the chamber
- Make some space to put the chamber (mind the grease on the bottom edge)
- Carefully lift the chamber. Good places to grab it are the tubing on the left and the right bottom edge.
Caution: The tube connecting the chamber with the pump will exert a force to the right!
Make sure you do not crash the chamber into the sample tray!
- Move the sample away from the gold mesh
- Move the tray fully down by turning the upper micrometer screw counterclockwise
- Move the tray fully to the left using the larger micrometer screw
- Loosen the screw on the clamp holding the sample in place and remove the sample
- Prepare the new sample
- Put the sample on the tray and place the clamp on it.
- Tighten the clamp screw (you might need to grab the nut on the bottom side)
- Move the tray in x-position using the large micrometer screw.
The gold mesh should be well aligned with the sample surface - Carefully move the tray up by turning the small micrometer screw clockwise.
Leave about half a millimeter of space between the sample and the gold mesh
Do not crash the sample into the gold mesh!
- Put the chamber back on
- Put the chamber back on and align it on the two posts on the bottom and right
- Check the backside to make sure that no cables are sticking out
- Vacuum the chamber
- Make sure the nitrogen venting valves are closed
- Make sure the turbopump is not spinning anymore
- When the turbopump has fully stopped, slowly open the valve on top of the pump
- Fully open the valve
- Wait until the vacuum is in the lower 1E-2 range, eg 4.0E-2 (read the value from the sensor attached to the turbopump)
- Only when this target pressure is reached, turn on the turbopump (control panel on top shelf)
- Wait until the desired vacuum for measurement is reached (usually 1E-5)