Added measurement methods that use overlappedY()
This commit is contained in:
parent
e1ba44127e
commit
a285fab1a1
@ -1,6 +1,6 @@
|
||||
"""
|
||||
run this before using this library:
|
||||
ipython -i keithley_interactive.py
|
||||
ipython -i k_teng_interactive.py
|
||||
|
||||
always records iv-t curves
|
||||
i-data -> smua.nvbuffer1
|
||||
@ -29,10 +29,10 @@ if __name__ == "__main__":
|
||||
|
||||
|
||||
from .keithley import keithley as _keithley
|
||||
from .keithley.measure import measure_count as _measure_count, measure as _measure
|
||||
from .utility import data as _data
|
||||
from .utility.data import load_dataframe
|
||||
from .utility import file_io
|
||||
from .utility import testing
|
||||
|
||||
_runtime_vars = {
|
||||
"last-measurement": ""
|
||||
@ -41,7 +41,7 @@ _runtime_vars = {
|
||||
settings = {
|
||||
"datadir": path.expanduser("~/data"),
|
||||
"name": "measurement",
|
||||
"interval": 0.01,
|
||||
"interval": 0.02,
|
||||
"beep": True,
|
||||
}
|
||||
config_path = path.expanduser("~/.config/k-teng.json")
|
||||
@ -52,85 +52,113 @@ test = False
|
||||
k = None
|
||||
|
||||
|
||||
def _measure(max_measurements=None, max_points_shown=None, monitor=False):
|
||||
|
||||
def _update_print(i, ival, vval):
|
||||
print(f"{i:5d} - {ival:.12f} A - {vval:.5f} V" + " "*10, end='\r')
|
||||
|
||||
class _Monitor:
|
||||
"""
|
||||
Monitor the voltage with matplotlib.
|
||||
Monitor v and i data
|
||||
"""
|
||||
def __init__(self, max_points_shown=None, use_print=False):
|
||||
self.max_points_shown = max_points_shown
|
||||
self.use_print = use_print
|
||||
self.index = []
|
||||
self.vdata = []
|
||||
self.idata = []
|
||||
|
||||
plt.ion()
|
||||
self.fig1, (self.vax, self.iax) = plt.subplots(2, 1)
|
||||
|
||||
self.vline, = self.vax.plot(self.index, self.vdata, color="g")
|
||||
self.vax.set_ylabel("Voltage [V]")
|
||||
self.vax.grid(True)
|
||||
|
||||
self.iline, = self.iax.plot(self.index, self.idata, color="m")
|
||||
self.iax.set_ylabel("Current [A]")
|
||||
self.iax.grid(True)
|
||||
|
||||
def update(self, i, ival, vval):
|
||||
if self.use_print:
|
||||
_update_print(i, ival, vval)
|
||||
self.index.append(i)
|
||||
self.idata.append(ival)
|
||||
self.vdata.append(vval)
|
||||
# update data
|
||||
self.iline.set_xdata(self.index)
|
||||
self.iline.set_ydata(self.idata)
|
||||
self.vline.set_xdata(self.index)
|
||||
self.vline.set_ydata(self.vdata)
|
||||
# recalculate limits and set them for the view
|
||||
self.iax.relim()
|
||||
self.vax.relim()
|
||||
if self.max_points_shown and i > self.max_points_shown:
|
||||
self.iax.set_xlim(i - self.max_points_shown, i)
|
||||
self.vax.set_xlim(i - self.max_points_shown, i)
|
||||
self.iax.autoscale_view()
|
||||
self.vax.autoscale_view()
|
||||
# update plot
|
||||
self.fig1.canvas.draw()
|
||||
self.fig1.canvas.flush_events()
|
||||
|
||||
def __del__(self):
|
||||
plt.close(self.fig1)
|
||||
|
||||
|
||||
def monitor_count(count=5000, interval=settings["interval"], max_points_shown=160):
|
||||
"""
|
||||
Take <count> measurements in <interval> and monitor live with matplotlib.
|
||||
|
||||
@details:
|
||||
- Resets the buffers
|
||||
- Opens a matplotlib window and takes measurements depending on settings["interval"]
|
||||
- Waits for the user to press a key
|
||||
@param max_points_shown : how many points should be shown at once. None means infinite
|
||||
@param max_measurements : maximum number of measurements. None means infinite
|
||||
You can take the data from the buffer afterwards, using save_csv """
|
||||
global k, settings, test, _runtime_vars
|
||||
print(f"Starting measurement with:\n\tinterval = {settings['interval']}s\nUse <C-c> to stop. Save the data using 'save_csv()' afterwards.")
|
||||
_runtime_vars["last_measurement"] = dtime.now().isoformat()
|
||||
if not test:
|
||||
_keithley.reset(k, verbose=True)
|
||||
k.write("smua.source.output = 1")
|
||||
k.write("format.data = format.ASCII\nformat.asciiprecision = 12")
|
||||
# jupyter:
|
||||
# clear_output(wait=True)
|
||||
# plt.plot(data)
|
||||
# plt.show()
|
||||
index = []
|
||||
vdata = []
|
||||
idata = []
|
||||
if monitor:
|
||||
plt.ion()
|
||||
fig1, (vax, iax) = plt.subplots(2, 1)
|
||||
Uses the device internal overlappedY measurement method, which allows for greater precision
|
||||
You can take the data from the buffer afterwards, using save_csv
|
||||
@param count: count
|
||||
@param interval: interval, defaults to settings["interval"]
|
||||
@param max_points_shown: how many points should be shown at once. None means infinite
|
||||
"""
|
||||
plt_monitor = _Monitor(max_points_shown, use_print=True)
|
||||
update_func = plt_monitor.update
|
||||
|
||||
vline, = vax.plot(index, vdata, color="g")
|
||||
vax.set_ylabel("Voltage [V]")
|
||||
vax.grid(True)
|
||||
|
||||
iline, = iax.plot(index, idata, color="m")
|
||||
iax.set_ylabel("Current [A]")
|
||||
iax.grid(True)
|
||||
print(f"Starting measurement with:\n\tinterval = {interval}s\nSave the data using 'save_csv()' afterwards.")
|
||||
try:
|
||||
i = 0
|
||||
while max_measurements is None or i < max_measurements:
|
||||
index.append(i)
|
||||
if test:
|
||||
idata.append(testing.testcurve(i, peak_width=1, amplitude=5e-8))
|
||||
vdata.append(-testing.testcurve(i, peak_width=2, amplitude=15))
|
||||
# data.append(np.random.rand())
|
||||
else:
|
||||
# data.append(float(k.query("print(smua.measure.v(smua.nvbuffer1))").strip('\n')))
|
||||
i_val, v_val = tuple(float(v) for v in k.query("print(smua.measure.iv(smua.nvbuffer1, smua.nvbuffer2))").strip('\n').split('\t'))
|
||||
idata.append(i_val)
|
||||
vdata.append(v_val)
|
||||
print(f"{i:5d} - {idata[-1]:.12f} A - {vdata[-1]:.5f} V", end='\r')
|
||||
if monitor:
|
||||
# update data
|
||||
iline.set_xdata(index)
|
||||
iline.set_ydata(idata)
|
||||
vline.set_xdata(index)
|
||||
vline.set_ydata(vdata)
|
||||
# recalculate limits and set them for the view
|
||||
iax.relim()
|
||||
vax.relim()
|
||||
if max_points_shown and i > max_points_shown:
|
||||
iax.set_xlim(i - max_points_shown, i)
|
||||
vax.set_xlim(i - max_points_shown, i)
|
||||
iax.autoscale_view()
|
||||
vax.autoscale_view()
|
||||
# update plot
|
||||
fig1.canvas.draw()
|
||||
fig1.canvas.flush_events()
|
||||
sleep(settings["interval"])
|
||||
i += 1
|
||||
_measure_count(k, V=True, I=True, count=count, interval=interval, beep_done=False, verbose=False, update_func=update_func, update_interval=0.05, testing=test)
|
||||
except KeyboardInterrupt:
|
||||
if not test:
|
||||
k.write("smua.source.output = 0")
|
||||
if monitor:
|
||||
plt.close(fig1)
|
||||
print("Measurement stopped" + " "*50)
|
||||
return vdata, idata
|
||||
k.write(f"smua.source.output = smua.OUTPUT_OFF")
|
||||
print("Monitoring cancelled, measurement might still continue" + " "*50)
|
||||
else:
|
||||
print("Measurement finished" + " "*50)
|
||||
|
||||
def measure_count(count=5000, interval=settings["interval"]):
|
||||
"""
|
||||
Take <count> measurements in <interval>
|
||||
|
||||
@details:
|
||||
- Resets the buffers
|
||||
- Takes <count> measurements depending on settings["interval"]
|
||||
Uses the device internal overlappedY measurement method, which allows for greater precision
|
||||
You can take the data from the buffer afterwards, using save_csv
|
||||
@param count: count
|
||||
@param interval: interval, defaults to settings["interval"]
|
||||
"""
|
||||
update_func = _update_print
|
||||
|
||||
print(f"Starting measurement with:\n\tinterval = {interval}s\nSave the data using 'save_csv()' afterwards.")
|
||||
try:
|
||||
_measure_count(k, V=True, I=True, count=count, interval=interval, beep_done=False, verbose=False, update_func=update_func, update_interval=0.05, testing=test)
|
||||
except KeyboardInterrupt:
|
||||
if not test:
|
||||
k.write(f"smua.source.output = smua.OUTPUT_OFF")
|
||||
print("Monitoring cancelled, measurement might still continue" + " "*50)
|
||||
else:
|
||||
print("Measurement finished" + " "*50)
|
||||
|
||||
|
||||
def monitor(max_measurements=None, max_points_shown=160):
|
||||
|
||||
|
||||
def monitor(interval=settings["interval"], max_measurements=None, max_points_shown=160):
|
||||
"""
|
||||
Monitor the voltage with matplotlib.
|
||||
|
||||
@ -138,13 +166,20 @@ def monitor(max_measurements=None, max_points_shown=160):
|
||||
- Resets the buffers
|
||||
- Opens a matplotlib window and takes measurements depending on settings["interval"]
|
||||
- Waits for the user to press a key
|
||||
Uses python's time.sleep() for waiting the interval, which is not very precise. Use measure_count for better precision.
|
||||
You can take the data from the buffer afterwards, using save_csv.
|
||||
@param max_points_shown : how many points should be shown at once. None means infinite
|
||||
@param max_measurements : maximum number of measurements. None means infinite
|
||||
You can take the data from the buffer afterwards, using save_csv """
|
||||
_measure(max_measurements=max_measurements, max_points_shown=max_points_shown, monitor=True)
|
||||
"""
|
||||
global _runtime_vars
|
||||
_runtime_vars["last_measurement"] = dtime.now().isoformat()
|
||||
print(f"Starting measurement with:\n\tinterval = {interval}s\nUse <C-c> to stop. Save the data using 'save_csv()' afterwards.")
|
||||
plt_monitor = _Monitor(use_print=True, max_points_shown=max_points_shown)
|
||||
update_func = plt_monitor.update
|
||||
_measure(k, interval=interval, max_measurements=max_measurements, update_func=update_func, testing=test)
|
||||
|
||||
|
||||
def measure(max_measurements=None):
|
||||
def measure(interval=settings["interval"], max_measurements=None):
|
||||
"""
|
||||
Measure voltages
|
||||
|
||||
@ -152,23 +187,45 @@ def measure(max_measurements=None):
|
||||
- Resets the buffers
|
||||
- Measure voltages
|
||||
- Waits for the user to press a key
|
||||
Uses python's time.sleep() for waiting the interval, which is not very precise. Use measure_count for better precision.
|
||||
You can take the data from the buffer afterwards, using save_csv.
|
||||
@param max_measurements : maximum number of measurements. None means infinite
|
||||
You can take the data from the buffer afterwards, using save_csv """
|
||||
_measure(max_measurements=max_measurements, monitor=False)
|
||||
"""
|
||||
global _runtime_vars
|
||||
_runtime_vars["last_measurement"] = dtime.now().isoformat()
|
||||
print(f"Starting measurement with:\n\tinterval = {interval}s\nUse <C-c> to stop. Save the data using 'save_csv()' afterwards.")
|
||||
update_func = _update_print
|
||||
_measure(k, interval=interval, max_measurements=max_measurements, update_func=update_func, testing=test)
|
||||
|
||||
def automeasure(repeat, repeat_delay=0, max_measurements=None, max_points_shown=120, monitor=True):
|
||||
|
||||
def repeat(measure_func: callable, count: int, repeat_delay=0):
|
||||
"""
|
||||
Measure and save to csv multiple times
|
||||
|
||||
@details
|
||||
Repeat count times:
|
||||
- call measure_func
|
||||
- call save_csv
|
||||
- sleep for repeat_delay
|
||||
|
||||
@param measure_func: The measurement function to use. Use a lambda to bind your parameters!
|
||||
@param count: Repeat count times
|
||||
|
||||
Example: Repeat 10 times:
|
||||
repeat(lambda : monitor_count(count=6000, interval=0.02, max_points_shown=200), 10)
|
||||
"""
|
||||
for i in range(repeat):
|
||||
_measure(max_measurements=max_measurements, max_points_shown=max_points_shown, monitor=monitor)
|
||||
try:
|
||||
for _ in range(count):
|
||||
measure_func()
|
||||
save_csv()
|
||||
sleep(repeat_delay)
|
||||
except KeyboardInterrupt:
|
||||
pass
|
||||
|
||||
|
||||
def get_dataframe():
|
||||
"""
|
||||
Get a pandas dataframe from the data in smua.nvbuffer1
|
||||
Get a pandas dataframe from the data in smua.nvbuffer1 and smua.nvbuffer2
|
||||
"""
|
||||
global k, settings, _runtime_vars
|
||||
if test:
|
||||
@ -241,16 +298,18 @@ def load_settings():
|
||||
global settings, config_path
|
||||
with open(config_path, "r") as file:
|
||||
settings = json.load(file)
|
||||
settings["datadir"] = path.expanduser(settings["datadir"])
|
||||
settings["datadir"] = path.expanduser(settings["datadir"]) # replace ~
|
||||
|
||||
def help(topic=None):
|
||||
if topic == None:
|
||||
print("""
|
||||
Functions:
|
||||
measure - measure the voltage
|
||||
monitor - measure the voltage with live monitoring in a matplotlib window
|
||||
automeasure - measure and save to csv multiple times
|
||||
get_dataframe - return smua.nvbuffer1 as pandas dataframe
|
||||
measure - take measurements
|
||||
monitor - take measurements with live monitoring in a matplotlib window
|
||||
measure_count - take a fixed number of measurements
|
||||
monitor_count - take a fixed number of measurements with live monitoring in a matplotlib window
|
||||
repeat - measure and save to csv multiple times
|
||||
get_dataframe - return smua.nvbuffer 1 and 2 as pandas dataframe
|
||||
save_csv - save the last measurement as csv file
|
||||
save_pickle - save the last measurement as pickled pandas dataframe
|
||||
load_dataframe - load a pandas dataframe from csv or pickle
|
||||
@ -303,7 +362,7 @@ def init():
|
||||
| < ______ | | | __)_ / | \ / \ ___
|
||||
| | \ /_____/ | | | \/ | \\ \_\ \
|
||||
|____|__ \ |____| /_______ /\____|__ / \______ /
|
||||
\/ \/ \/ \/ 1.0
|
||||
\/ \/ \/ \/ 1.1
|
||||
Interactive Shell for TENG measurements with Keithley 2600B
|
||||
---
|
||||
Enter 'help()' for a list of commands""")
|
||||
|
@ -1,115 +1,106 @@
|
||||
from time import sleep
|
||||
import numpy as np
|
||||
from matplotlib import pyplot as plt
|
||||
import pyvisa
|
||||
|
||||
smua_settings = """
|
||||
display.clear()
|
||||
display.settext('starting')
|
||||
smua.reset()
|
||||
smua.measure.autorangev = smua.AUTORANGE_ON
|
||||
smua.measure.autozero = smua.AUTOZERO_ONCE
|
||||
smua.source.output = smua.OUTPUT_OFF
|
||||
-- max 20 V expected
|
||||
smua.measure.rangev = 20
|
||||
from .keithley import reset
|
||||
from ..utility import testing as _testing
|
||||
|
||||
"""
|
||||
|
||||
script_dir = "scripts/"
|
||||
scripts = {
|
||||
"buffer_reset": "buffer_reset.lua",
|
||||
"smua_reset": "smua_reset.lua",
|
||||
}
|
||||
for key,val in scripts.items():
|
||||
scripts[key] = script_dir + scripts[key]
|
||||
|
||||
|
||||
def run_lua(instr, script_path, verbose=False):
|
||||
def measure_count(instr, V=True, I=True, count=100, interval=0.05, update_func=None, update_interval=0.5, beep_done=True, verbose=True, testing=False):
|
||||
"""
|
||||
Run a lua script from the host on the instrument
|
||||
@param instr : pyvisa instrument
|
||||
@param script_path : full path to the script
|
||||
"""
|
||||
with open(script_path, "r") as file:
|
||||
script = file.read()
|
||||
if verbose: print(f"Running script: {script_path}")
|
||||
instr.write(script)
|
||||
Take <count> measurements with <interval> inbetween
|
||||
|
||||
|
||||
def measure_V(instr, count=100, interval=0.05):
|
||||
@details
|
||||
Uses the devices overlappedY function to make the measurements asynchronosly
|
||||
The update_func is optional and only used when I == True and V == True
|
||||
The update_func does not necessarily get all the values that are measured. To obtain the whole measurement, get them from the device buffers (smua.nvbufferX)
|
||||
@param instr: pyvisa instrument
|
||||
@param update_func: Callable that processes the measurements: (index, ival, vval) -> None
|
||||
@param update_interval: interval at which the update_func is called
|
||||
"""
|
||||
"""
|
||||
data = []
|
||||
for _ in range(1000):
|
||||
data.append(tuple(float(v) for v in instr.query("print(smua.measure.v())").strip('\n').split('\t')))
|
||||
# print(i, data[-1])
|
||||
# clear_output(wait=True)
|
||||
plt.plot(data)
|
||||
plt.show()
|
||||
sleep(0.05)
|
||||
|
||||
def reset(instr, verbose=False):
|
||||
"""
|
||||
Reset smua and its buffers
|
||||
@param instr : pyvisa instrument
|
||||
"""
|
||||
run_lua(instr, scripts["smua_reset"], verbose=verbose)
|
||||
run_lua(instr, scripts["buffer_reset"], verbose=verbose)
|
||||
|
||||
|
||||
def measure_count(instr, V=True, I=True, count=100, interval=0.05, beep_done=True, verbose=True):
|
||||
"""
|
||||
take n measurements at dt interval
|
||||
@param instr : pyvisa instrument
|
||||
"""
|
||||
reset(instr, verbose=verbose)
|
||||
f_meas = None
|
||||
if V and I:
|
||||
f_meas = "smua.measure.iv(smua.nvbuffer1, smua.nvbuffer2)"
|
||||
f_meas = "smua.measure.overlappediv(smua.nvbuffer1, smua.nvbuffer2)"
|
||||
elif V:
|
||||
f_meas = "smua.measure.v(smua.nvbuffer1)"
|
||||
f_meas = "smua.measure.overlappedv(smua.nvbuffer1)"
|
||||
elif I:
|
||||
f_meas = "smua.measure.i(smua.nvbuffer1)"
|
||||
f_meas = "smua.measure.overlappedi(smua.nvbuffer1)"
|
||||
else:
|
||||
print("I and/or V needs to be set to True")
|
||||
return
|
||||
|
||||
i = 0
|
||||
if not testing:
|
||||
reset(instr, verbose=verbose)
|
||||
instr.write(f"smua.measure.count = {count}")
|
||||
instr.write(f"smua.measure.interval = {interval}")
|
||||
|
||||
# start measurement
|
||||
instr.write(f"smua.source.output = smua.OUTPUT_ON")
|
||||
instr.write(f_meas)
|
||||
|
||||
condition = lambda: float(instr.query("print(status.operation.measuring.condition)").strip("\n ")) != 0
|
||||
else:
|
||||
condition = lambda: i < int(float(count) * interval / update_interval)
|
||||
|
||||
sleep(update_interval)
|
||||
# for live viewing
|
||||
|
||||
# will return 2.0 while measruing
|
||||
while condition():
|
||||
if update_func and V and I:
|
||||
try:
|
||||
if not testing:
|
||||
ival = float(instr.query("print(smua.nvbuffer1.readings[smua.nvbuffer1.n])").strip("\n"))
|
||||
vval = float(instr.query("print(smua.nvbuffer2.readings[smua.nvbuffer2.n])").strip("\n"))
|
||||
else:
|
||||
ival = _testing.testcurve(i, peak_width=1, amplitude=5e-8)
|
||||
vval = -_testing.testcurve(i, peak_width=2, amplitude=15)
|
||||
update_func(i, ival, vval)
|
||||
except ValueError:
|
||||
pass
|
||||
sleep(update_interval)
|
||||
i += 1
|
||||
|
||||
if not testing:
|
||||
instr.write(f"smua.source.output = smua.OUTPUT_OFF")
|
||||
|
||||
if beep_done:
|
||||
instr.write("beeper.beep(0.3, 1000)")
|
||||
|
||||
|
||||
|
||||
def event_test_TODO():
|
||||
# Type of event we want to be notified about
|
||||
event_type = pyvisa.constants.EventType.service_request
|
||||
# Mechanism by which we want to be notified
|
||||
event_mech = pyvisa.constants.EventMechanism.queue
|
||||
keithley.enable_event(event_type, event_mech)
|
||||
|
||||
# Instrument specific code to enable service request
|
||||
# (for example on operation complete OPC)
|
||||
keithley.write("*SRE 1")
|
||||
keithley.write("INIT")
|
||||
|
||||
|
||||
with open("script.lua", "r") as file:
|
||||
script = file.read()
|
||||
# for line in script.split('\n'):
|
||||
# input(line)
|
||||
# keithley.write(line)
|
||||
keithley.write(script)
|
||||
|
||||
# Wait for the event to occur
|
||||
response = keithley.wait_on_event(event_type, 1000)
|
||||
assert response.event.event_type == event_type
|
||||
assert response.timed_out == False
|
||||
|
||||
instr.disable_event(event_type, event_mech)
|
||||
keithley.query_ascii_values("printbuffer(1, 10, smua.nvbuffer1)", 6)
|
||||
print(voltages)
|
||||
def measure(instr, interval, update_func=None, max_measurements=None, testing=False):
|
||||
"""
|
||||
@details:
|
||||
- Resets the buffers
|
||||
- Until KeyboardInterrupt:
|
||||
- Take measurement
|
||||
- Call update_func
|
||||
- Wait interval
|
||||
Uses python's time.sleep() for waiting the interval, which is not very precise. Use measure_count for better precision
|
||||
You can take the data from the buffer afterwards, using save_csv
|
||||
@param instr: pyvisa instrument
|
||||
@param update_func: Callable that processes the measurements: (index, ival, vval) -> None
|
||||
@param max_measurements : maximum number of measurements. None means infinite
|
||||
"""
|
||||
if not testing:
|
||||
reset(instr, verbose=True)
|
||||
instr.write("smua.source.output = smua.OUTPUT_ON")
|
||||
instr.write("format.data = format.ASCII\nformat.asciiprecision = 12")
|
||||
try:
|
||||
i = 0
|
||||
while max_measurements is None or i < max_measurements:
|
||||
if testing:
|
||||
ival = _testing.testcurve(i, peak_width=1, amplitude=5e-8)
|
||||
vval = -_testing.testcurve(i, peak_width=2, amplitude=15)
|
||||
else:
|
||||
ival, vval = tuple(float(v) for v in instr.query("print(smua.measure.iv(smua.nvbuffer1, smua.nvbuffer2))").strip('\n').split('\t'))
|
||||
if update_func:
|
||||
update_func(i, ival, vval)
|
||||
sleep(interval)
|
||||
i += 1
|
||||
except KeyboardInterrupt:
|
||||
pass
|
||||
if not testing:
|
||||
instr.write("smua.source.output = smua.OUTPUT_OFF")
|
||||
print("Measurement stopped" + " "*50)
|
||||
|
Loading…
Reference in New Issue